17-18
Model 2790 SourceMeter
®
Switch System Reference Manual
DC current calibration
After the 1M
Ω
resistance point has been calibrated, follow these steps for DC current
calibration:
1.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2790
as shown in
2.
Perform the calibration steps listed in
. For each step:
•
Set the calibrator to the indicated current and make sure the unit is in operate.
(Use the recommended current if possible.)
•
Send the indicated programming command. (Change the current parameter if
you are using a different calibration current.)
•
Wait until the Model 2790 completes each step before continuing.
klqb
If you are performing DC-only calibration, proceed to “Programming calibra-
tion dates.”
AC voltage calibration
Follow these steps for AC voltage calibration:
1.
Connect the calibrator to the Model 2790 INPUT HI and LO terminals as shown in
2.
Perform the calibration steps summarized in
. For each step:
•
Set the calibrator to the indicated voltage and frequency. Make sure the unit is
in operate. (You must use the stated voltage and frequency.)
•
Send the indicated programming command.
•
Wait until the Model 2790 completes each step before continuing.
Table 17-10
DC current calibration programming steps
Calibration
step
Calibrator
current
Calibration command*
Parameter range
10mA
100mA
1A
10.00000mA
100.00000mA
1.000000A
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1
*Change parameter if using different current.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...