Resistance calibration
Completing the 100V DC calibration step ends the DC voltage calibration procedure. The
Model 2790 will then prompt you to connect 1k
Ω
. Follow these steps for resistance cali-
bration:
1.
Set the calibrator output for resistance, and turn on external sense.
klqb
Use external sense (4-wire
Ω
) when calibrating all resistance ranges. Be sure
that the calibrator external sense mode is turned on.
2.
Perform the calibration steps summarized in
. For each step:
•
Set the calibrator to the indicated value, and place the unit in operate. (If the
calibrator cannot output the exact resistance value, use the Model 2790 left
and right arrow keys and the range keys to adjust the Model 2790 display to
agree with the actual calibrator resistance.)
•
Press the ENTER key to calibrate each point.
•
Wait for the Model 2790 to complete each step before continuing.
DC current calibration
After the 1M
Ω
resistance point has been calibrated, the unit will prompt you to apply
10mA. Follow these steps for DC current calibration:
1.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2790
as shown in
Table 17-4
Resistance calibration summary
Calibration
step
Calibrator
resistance*
Allowable
range
1k
Ω
10k
Ω
100k
Ω
1M
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
0.9k
Ω
to 1.1k
Ω
9k
Ω
to 11k
Ω
90k
Ω
to 110k
Ω
0.9M
Ω
to 1.1M
Ω
* Nominal resistance. Adjust Model 2790 calibration parameter to
agree with actual value.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...