16-36
Model 2790 SourceMeter
®
Switch System Reference Manual
3.
Set the Model 7751/7753 source voltage to 500V as follows:
a.
Press SHIFT, CARD, then CONFIG.
b.
Select SLOT1: 7751 or SLOT1: 7753, then press ENTER.
c.
Select the V-C submenu, then set the source to 500V.
d.
Press EXIT to return to normal display.
4.
Using the CLOSE key, close the following slot 2 channels: 1, 2, 16, 18, 21, 22,
and 23.
5.
Set the math function to display resistance as follows:
a.
Press SHIFT then MATH.
b.
Select S1VOHMS, then press ENTER.
c.
Press EXIT to return to normal display.
6.
Recalculate reading limits shown in
(7751) or
(7753)
based on the actual characterized resistance value.
7.
Verify that the resistance reading is within the limits calculated in step 6.
8.
Repeat steps 1 through 7 for the remaining resistances and voltage source settings
as shown in
(7751) or
(7753).
Table 16-16
Limits for Model 7751 source-voltage resistance verification
Source
current
Nominal
resistance
Nominal reading limits
(1 year, 18°C to 28°C)
Recalculated limits*
500V
500V
50V
50V
1G
Ω
100M
Ω
100M
Ω
10M
Ω
0.98186 to 1.01814G
Ω
98.585 to 101.415M
Ω
94.048 to 105.952M
Ω
9.4747 to 10.5253M
Ω
__________ to __________
Ω
__________ to __________
Ω
__________ to __________
Ω
__________ to __________
Ω
* Calculate limits based on actual characterized resistance values and Model 7751 resistance accuracy
specifications. See
Verification limits
.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...