Model 2790 SourceMeter
®
Switch System Reference Manual
7-39
Table 7-3
Monitor scan example
Front panel operation
Remote programming
1
Restore defaults (SHIFT SETUP > RESTORE: FACT).
SYST:PRES
2
For front panel operation, proceed to step 3.
For remote programming, clear the buffer.
TRAC:CLE
3
Configure advanced scan:
SHIFT CONFIG > ADVANCED:
a
Channel 101:
Select TEMP function.
FUNC 'TEMP',(@101)
Configure temperature (SHIFT SENSOR):
Select thermistor sensor (SENS: THRMSTR).
TEMP:TRAN THER,(@101)
Select thermistor type (TYPE: 5000
Ω
).
TEMP:THER 5000,(@101)
b
Set and enable high limit 1:
Set limit to 30 (SHIFT LIMITS > HI1:+30.00000).
CALC3:LIM1:UPP 30,(@101)
Enable (on) limit (SHIFT OFF/ON > LIMITS: ON).
CALC3:LIM1:STAT ON,(@101)
Enable Channel Average (SHIFT CH AVG).
CAV ON,(@101)
Close channel 1.
ROUT:CLOS (@101)
c
Channel 102, 103, and 104:
Select DCV function.
FUNC 'VOLT',(@102:104)
Select 10V range.
VOLT:RANG 10,(@102:104)
Set filter count to 20 (SHIFT TYPE > 020 RDGS).
VOLT:AVER:COUN 20,(@102:104)
Enable filter (FILTER).
VOLT:AVER:STAT ON,(@102:104)
d
Disable (off) channels 105 through 222
(SHIFT CH OFF).
ROUT:SCAN (@101:104)
e
Disable immediate scan (IMM SCAN: N) and enable
high limit 1 (HLIM1 SCAN:Y).
ROUT:SCAN:TSO HLIM1
f
Disable timer (TIMER? NO).
g
Set reading count to 4.
ROUT:MON:POIN 4
4
Select and enable monitor channel
(SHIFT MONITOR >101).
ROUT:MON (@101)
ROUT:MON:STAT ON
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...