Model 2790 SourceMeter
®
Switch System Reference Manual
7-29
Scanning commands
Scanning commands are listed in
. Additional information on these commands
follow the table.
klqb
Memory pattern commands, queries, and optional command words are not
included in
. The unabridged SCPI tables are provided in
Memory pattern commands are listed in
and in
Table 7-1
Scanning commands
Commands
Description
Default Ref
Scan commands
ROUTe:SCAN <clist>
Specify list of channels to be scanned.
a
ROUTe:SCAN?
Returns list of channels to be scanned.
ROUTe:SCAN:TSOurce <list>
Select trigger(s) to start scan: <list> = IMMediate, or
HLIMit1, LLIMit1, HLIMit2, LLIMit2.
IMM
b
ROUTe:SCAN:NVOLatile <b>
Enable or disable auto scan; <b> = ON or OFF.
(Note 1)
ROUTe:SCAN:LSELect <name> Enable/disable scan; <name> = INTernal (on) or NONE
(off).
NONE
ROUTe:MONitor <clist>
Specify one channel to be monitored.
c
ROUTe:MONitor:POINts <NRf> Specify number of channels to scan; <NRf> = 2 to
55000.
(Note 3)
d
ROUTe:MONitor:STATe <b>
Enable/disable monitor; <b> = ON or OFF.
OFF
ROUTe:MONitor:DATA?
Returns the most recent monitor reading.
ROUTe:CLOSe:ACONfigure <b> Enable or disable auto channel configuration.
Limits commands
CALCulate3:LIMit1:UPPer <n>
Set high limit 1 for monitor; <n> = -4294967295 to
+4294967295.
1.0
CALCulate3:LIMit1:LOWer <n> Set low limit 1 for monitor; <n> = -4294967295 to
+4294967295.
-1.0
CALCulate3:LIMit2:UPPer <n>
Set high limit 2 for monitor; <n> = -4294967295 to
+4294967295.
2.0
CALCulate3:LIMit2:LOWer <n> Set low limit 2 for monitor; <n> = -4294967295 to
+4294967295.
-2.0
CALCulate3:LIMit1:STATe <b>
Enable/disable limit 1 test; <b> = ON or OFF.
OFF
CALCulate3:LIMit2:STATe <b>
Enable/disable limit 2 test; <b> = ON or OFF.
OFF
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...