Range
The range setting is “remembered” by each measurement function. When you select a
function, the instrument will return to the last range setting for that function.
Measurement ranges and maximum readings
The selected range affects both accuracy of the measurement as well as the maximum
level that can be measured. The measurement ranges and maximum readings for all func-
tions, except FREQ, PERIOD, and TEMP, are listed in
.
Input values that exceed the maximum readings cause the message “OVERFLOW” to be
displayed.
FREQ and PERIOD
— Frequency measurements from 3Hz to 500kHz and period
measurements from 2µs to 333µs can be made on the ACV ranges.
TEMP
— There is no range selection for temperature measurements. Temperature
measurements are performed on a single, fixed range. Depending on which type of sensor
is being used, the maximum temperature readings range from –200°C to 630°C.
,” lists the reading range for each sensor type.
Table 4-1
Measurement ranges and maximum readings
Function
Ranges
Maximum
reading
DCV
100mV, 1V, 10V, 100V, 1000V
±1010V
ACV
100mV, 1V, 10V, 100V, 750V
757.5V
DCI
20mA, 100mA, 1A, 3A
±3.1A
ACI
1A, 3A
3.1A
Ω
2
100
Ω
, 1k
Ω
, 10k
Ω
, 100k
Ω
, 1M
Ω
, 10M
Ω
, 100M
Ω
120M
Ω
Ω
4
*
100
Ω
, 1k
Ω
, 10k
Ω
, 100k
Ω
, 1M
Ω
, 10M
Ω
, 100M
Ω
120M
Ω
*Offset compensated ohms (OCOMP) can be performed on the 100¾, 1k¾, and 10k¾ ranges.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...