Model 2750 Multimeter/Switch System User’s Manual
Signal Processing Sequence and Data Flow
D-11
CALC[1]:DATA[:LATest]?
CALC[1]:DATA:FRESh?
These two commands are similar to the LATest? and FRESh? commands for the SENSe1
subsystem, except that returned data arrays are the result of the math calculation. See
[SENS[1]]:DATA[:LATest]? and [SENS[1]]:DATA:FRESh? for details on the differences
between LATest? and FRESh?.
With a Math function enabled, both CALC[1]:DATA? and CALC[1]:DATA:FRESh?
return a single data array whose reading is the result of the math calculation. Note that the
calculation is performed on the last data array stored in the sample buffer. These com-
mands do not affect data in the sample buffer. Therefore, subsequent executions of these
commands return the same data.
With no Math function enabled, these commands return the last data array stored in the
sample buffer.
CALC3:LIM1:FAIL?
CALC3:LIM2:FAIL?
Each reading applied to the CALC 3 Limit Tests block is tested when Limits operations
are enabled. When comparing the reading to the programmed high and low limits of
Limit 1, CALC3:LIM1:FAIL? returns a “0” (inside limits) or a “1” (outside limits). Simi-
larly, CALC3:LIM2:FAIL? compares the reading to the high and low limits of Limit 2.
NOTE
Each data array returned by the read commands (SENS:DATA?, FETCh?,
READ?, MEAS?, CALC2:DATA?, TRACe:DATA?, and CALC1:DATA?) will
include the result code for limit tests if the limits data element is selected. See
“FORMat:ELEMents <item list>,” page 14-5
, to select the limits element and
interpret the code.
TRACe:DATA?
When the data store is enabled, sample buffer data or CALC1 results are stored in the
buffer. The TRACe:DATA? command reads the entire contents of the data store. The
selected feed (TRAC:FEED SENSe or TRAC:FEED CALC1), determines which group of
readings are stored.
2750-900-01.book Page 11 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...