Model 2750 Multimeter/Switch System User’s Manual
FORMat and Misc SYSTem Commands
14-5
FORMat:ELEMents <item list>
Parameters
READing
= DMM reading
UNITs
= Units
TSTamp
= Timestamp
RNUMber
= Reading number
CHANnel
= Channel number
LIMits
= Limits reading
The specified elements are included in the data string in response to FETCh?, READ?,
MEASure?, and TRACe:DATA?. Note that each element in the item list must be separated
by a comma (i.e. send “FORMat:ELEMents READing, UNITs, TSTamp, RNUMber,
CHANnel, LIMits”) to include all elements in the data string. The elements for the ASCII
format are shown in
.
An overflow reading is returned as +9.9E37. When a specified data element has invalid
data associated with it, NAN (Not A Number) will be the response. NAN is returned as
+9.9E37.
Timestamp
— Timestamp references the returned data string to a point in time. There are
two timestamps: relative and real-time clock. The following command selects the
timestamp:
SYSTem:TSTamp:TYPE <name> ' Select timestamp type; RELative or RTCL.
NOTE
The real-time clock timestamp can only be returned for the ASCII data format.
For the binary formats, timestamp will not be sent with the real-time clock
selected.
The relative timestamp operates as a timer that starts at zero seconds when the instrument
is turned on, or when the relative timestamp is reset (SYSTem:TSTamp:RELative:RESet).
After 99,999.99 seconds, the timer resets back to zero and starts over.
For buffer readings recalled from the front panel, the relative timestamp is referenced to
the first reading stored in the buffer (absolute format) which is timestamped at 0 seconds,
and to the time between each stored reading (delta format). For remote programming, you
can only return the absolute or delta timestamp. The following command is used to select
relative timestamp format for the buffer:
TRACe:TSTamp:FORMat ' Select timestamp format; ABSolute or DELTa.
Reading number
— The reading counter starts at zero when the Model 2750 is turned on.
When returning buffer readings using TRACe:DATA?, each reading will be referenced to
the first reading, which is #0. The following command will reset the counter:
SYSTem:RNUMber:RESet.
Channel number
— The channel number indicates the switching module channel for the
reading. Channel number 000 indicates that no channel was closed.
2750-900-01.book Page 5 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...