Model 2750 Multimeter/Switch System User’s Manual
Triggering
8-7
External trigger
The EXT TRIG input requires a falling-edge, TTL-compatible pulse with the specifica-
tions shown in
. In general, external triggers can be used to control measure
operations. For the Model 2750 to respond to external triggers, the trigger model must be
configured for it.
Figure 8-4
Trigger link input pulse specifications (EXT TRIG)
Voltmeter complete
The VMC output provides a TTL-compatible output pulse that can be used to trigger other
instruments. The specifications for this trigger pulse are shown in
. Typically,
you would want the Model 2750 to output a trigger after the settling time of each
measurement.
Figure 8-5
Trigger link output pulse specifications (VMC)
Triggers on
Leading Edge
TTL High
(2V-5V)
TTL Low
(
≤
0.8V)
2
μ
s
Minimum
Meter
Complete
TTL High
(3.4V Typical)
TTL Low
(0.25V Typical)
10
μ
s
Minimum
2750-900-01.book Page 7 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...