Model 2750 Multimeter/Switch System Service Manual
Performance Verification
1-35
770
0 V
eri
fication
Verifying ratio and average
Follow the procedure below to verify ratio and average.
CAUTION
Exceeding 300V between the 7700 plug-in module INPUT or SENSE H
and L terminals may cause instrument damage.
1.
Connect the Model 7700 CH1 and CH11 H and L terminals to the DC calibrator, as
shown in
.
2.
Install the Model 7700 in Slot 1 of the Model 2750, then turn on the power and
allow the unit to warm up for two hours before proceeding. Be sure the front panel
INPUTS switch is set to the REAR position.
3.
Select the Model 2750 DCV function and the 1V range. Close Channel 1 by press-
ing the CLOSE key, then ENTER for “SINGLE” channel, and keying in 101.
4.
Select the Model 2750 RATIO function (press SHIFT then RATIO).
5.
Set the calibrator output to 1.00000V DC, and allow the reading to settle.
6.
Verify that the ratio reading is between 0.9999926 and 1.000074.
7.
Press OPEN to open Channel 1.
Figure 1-18
Connections for Model 7700 ratio and average verification
H L H L
AMPS
H L H L
H L H L
H L H L
LO
CH21 CH22
CH11 CH12 CH13 CH14 CH15 CH16
H L H L H L H L
CH17 CH18 CH19 CH20
SENSE
(OHMS, 4 WIRE)
INPUT
(V, 2 WIRE)
H L H L H L H L
CH7
CH8 CH9 CH10
H L H L H L H L
H L H L H L H L
INPUT SENSE CH1
CH2
CH3
CH4
CH5
CH6
CH1
CH11
Output
HI
Output
LO
DC Voltage Calibrator
Note: Use shielded cables to
minimize noise.
Model 7700
Summary of Contents for 2750
Page 13: ...1 PerformanceVerification...
Page 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Page 49: ...2 Calibration...
Page 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Page 75: ...Routine Maintenance 3 RoutineMaintenance...
Page 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Page 87: ...Troubleshooting 4 Troubleshooting...
Page 111: ...Disassembly 5 Disassembly...
Page 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Page 119: ...Replaceable Parts 6 ReplaceableParts...
Page 135: ...Specifications A AccuracyandOptimization...
Page 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Page 141: ...Calibration Reference B CalibrationReference...
Page 167: ...Calibration Program C CalibrationProgram...
Page 178: ......
Page 179: ......