14-2
FORMat and Misc SYSTem Commands
Model 2700 Multimeter/Switch System User’s Manual
FORMat commands
The commands in this subsystem are used to select the format for transferring data,
, over the bus.
FORMat[:DATA] <type>[,<length>]
Parameters
ASCii
= ASCII format
SREal
= Binary IEEE-754 single precision format
REAL, 32
= Binary IEEE-754 single precision format
DREal
= Binary IEEE-754 double precision format
REAL, 64
= Binary IEEE-754 double precision format
NOTE
<length> is not used for the ASCii, SREal, or DREal parameters.
The response to READ?, FETCh?, MEASure?, TRACe:DATA?, CALC1:DATA?, or
CALC2:DATA? over the GPIB can be returned in either the ASCii or binary format. All
other queries are returned in ASCii, regardless of the selected format. Over the RS-232
interface, only the ASCII format is allowed.
NOTE
Regardless of which data format for output strings is selected, the instrument
will only respond to input commands using the ASCII format.
Table 14-1
SCPI commands — data format
Command
Description
Default
FORMat[:DATA] <type>[,<length>]
Specify data format; ASCii, SREal, REAL, 32,
DREal, or REAL, 64.
ASCii
FORMat:ELEMents <item list>
Specify data elements; READing, UNITs, TSTamp,
RNUMber, CHANnel, or LIMits.
All 3
FORMat:BORDer <name>
Specify byte order; NORMal or SWAPped.
(see Note)
Note: *RST default is NORMal. SYSTem:PRESet default is SWAPped.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......