Model 2700 Multimeter/Switch System User’s Manual
Status Structure
11-19
Event enable registers
shows, each status register set has an enable register. Each event register
bit is logically ANDed (&) to a corresponding enable bit of an enable register. Therefore,
when an event bit is set and the corresponding enable bit is set (as programmed by the
user), the output (summary) of the register will set to 1, which in turn sets the summary bit
of the Status Byte Register.
The commands to program and read the event enable registers are listed in
. For
details on programming and reading registers, see
“Programming enable registers,”
“Reading registers,” page 11-6
.
NOTE
The bits of any enable register can be reset to 0 by sending the 0
parameter value with the appropriate enable command (i.e.
STATus:OPERation:ENABle 0).
Table 11-5
Event enable registers commands
Command
Description
*ESE <NRf>
Program Standard Event Enable Register (0 to 255).
*ESE?
Read Standard Event Enable Register.
STATus:OPERation:ENABle <NRF>
Program Operation Event Enable Register (0 to 32767).
STATus:OPERation:ENABle?
Read enable register.
STATus:MEASurement:ENABle <NRf>
Program Measurement Event Enable Register (0 to 32767).
STATus:MEASurement:ENABle?
Read enable register.
STATus:QUEStionable:ENABle <NRf>
Program Questionable Event Enable Register (0 to 32767).
STATus:QUEStionable:ENABle?
Read enable register.
Note: Power-up and STATus:PRESet resets all bits of all enable registers to 0. *CLS has no effect.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......