Model 2700 Multimeter/Switch System User’s Manual
Status Structure
11-9
Status byte and service request commands
The commands to program and read the Status Byte Register and Service Request Enable
Register are listed in
. For details on programming and reading registers, see
“Programming enable registers,” page 11-5
“Reading registers,” page 11-6
.
NOTE
To reset the bits of the Service Request Enable Register to 0, use 0 as the
parameter value for the *SRE command (*SRE 0).
Programming example — set MSS (B6) when error occurs
The second command in the following sequence enables EAV (error available). When an
invalid command is sent (line 3), bits B2 (EAV) and B6 (MSS) of the Status Byte Register
set to 1. The last command reads the Status Byte Register. Keep in mind that you have to
address the Model 2700 to talk after sending a query command. To determine the exact
nature of the error, you will have to read the Error Queue (see
).
NOTE
The following example can be run from the KE2700 Instrument Driver using the
example named “PollSRQ” in
of
*CLS
' Clear Error Queue
*SRE 4
' Enable EAV.
*XYZ
' Generate error.
*STB?
' Read Status Byte Register.
Table 11-2
Status byte and service request enable register commands
Command
Description
*STB?
Read Status Byte Register.
*SRE <NRf>
Program the Service Request Enable Register (0 to 255).
*SRE?
Read the Service Request Enable Register.
Note: *CLS and STATus:PRESet have no effect on the Service Request Enable Register.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......