8-4
Triggering
Model 2700 Multimeter/Switch System User’s Manual
Delay (auto or manual)
A programmable delay is available after event detection. It can be set manually or an auto
delay can be used. With auto delay selected, the instrument automatically selects a delay
period that will provide sufficient settling for function and autorange changes and multi-
phase measurements.
Normal measurement state
— With auto delay selected and the External or Bus control
source selected, the Model 2700 selects a delay based on the selected voltage range. The
auto delay period cannot be adjusted by the user. The auto delays are listed in
.
With one of the other control sources selected, the auto delay is 0.000s for all functions
and ranges.
Scanning
— When scanning, the nominal delay will be long enough to allow each switch-
ing module channel relay to settle before making the measurement. When scanning, the
auto delay times in
are valid for all control sources.
The delay function is accessed by pressing SHIFT and then DELAY. The present delay
setting (AUTO or MANual) is displayed. Press the
Δ
or
∇
key to display the desired
setting and press ENTER.
Table 8-1
Auto delay settings
Function
Range and delay
DCV
100mV
1ms
1V
1ms
10V
1ms
100V
5ms
1000V
5ms
ACV
100mV
400ms
1V
400ms
10V
400ms
100V
400ms
750V
400ms
FREQ and
PERIOD
100mV
1ms
1V
1ms
10V
1ms
100V
1ms
750V
1ms
DCI
20mA
2ms
100mA
2ms
1A
2ms
3A
2ms
ACI
1A
400ms
3A
400ms
Ω
2,
Ω
4
100
Ω
3ms
1k
Ω
3ms
10k
Ω
13ms
100k
Ω
25ms
1M
Ω
100ms
10M
Ω
150ms
100M
Ω
250ms
Continuity
1k
Ω
3ms
TEMP
The auto delay for thermocouples is 1ms. For thermistors and 4-wire
RTDs, the auto delay period is the same as the delay for the resistance
range that is used for the measurement.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......