4-2
Range, Digits, Rate, Bandwidth, and Filter
Model 2700 Multimeter/Switch System User’s Manual
Range
The range setting is “remembered” by each measurement function. When you select a
function, the instrument will return to the last range setting for that function.
Measurement ranges and maximum readings
The selected range affects both accuracy of the measurement as well as the maximum
level that can be measured. The measurement ranges and maximum readings for all
functions, except FREQ, PERIOD, and TEMP, are listed in
Input values that exceed the maximum readings cause the message “OVERFLOW” to be
displayed.
FREQ and PERIOD
— Frequency measurements from 3Hz to 500kHz, and period
measurements from 2µs to 333µs can be made on the ACV ranges.
TEMP
— There is no range selection for temperature measurements. Temperature
measurements are performed on a single, fixed range. Depending on which type of sensor
is being used, the maximum temperature readings range from –200°C to 1820°C.
,” lists the reading range for each sensor type.
Table 4-1
Measurement ranges and maximum readings
Function
Ranges
Maximum
reading
DCV
100mV, 1V, 10V, 100V, 1000V
±1010V
ACV
100mV, 1V, 10V, 100V, 750V
757.5V
DCI
20mA, 100mA, 1A, 3A
±3.1A
ACI
1A, 3A
3.1A
Ω
2
100
Ω
, 1k
Ω
, 10k
Ω
, 100k
Ω
, 1M
Ω
, 10M
Ω
, 100M
Ω
120M
Ω
Ω
4
*
100
Ω
, 1k
Ω
, 10k
Ω
, 100k
Ω
, 1M
Ω
, 10M
Ω
, 100M
Ω
120M
Ω
*Offset compensated ohms (OCOMP) can be performed on the 100
Ω
, 1k
Ω,
and 10k
Ω
ranges.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......