Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-220
2657A-901-01 Rev. B/December 2012
Also see
(on page 7-212)
smuX.source.offfunc
This attribute sets the source function that is used (source 0 A or 0 V) when the output is turned off and the
source-measure unit (SMU) is in normal output-off mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
1 (smuX.OUTPUT_DCVOLTS)
Usage
offfunc
= smu
X
.source.offfunc
smu
X
.source.offfunc =
offfunc
offfunc
Set to the source function to be used when the output is off and the SMU is in
normal output-off mode. Set to one of the following values:
0
or
smu
X
.OUTPUT_DCAMPS
: Source 0 A
1
or
smu
X
.OUTPUT_DCVOLTS
: Source 0 V
X
SMU channel (for example,
smua.source.offfunc
applies to SMU channel A)
Details
This attribute controls the source function used when the output is turned off and the
smu
X
.source.offmode
is
set to
smu
X
.OUTPUT_NORMAL
. Set this attribute to
smu
X
.OUTPUT_DCVOLTS
for the source to be a 0 V source
when the output is off (
smu
X
.source.offlimiti
is used). Set it to
smu
X
.OUTPUT_DCAMPS
for the source to
be a 0 A source when the output is off (
smu
X
.source.offlimitv
is used).
This attribute is only used when the
smu
X
.source.offmode
attribute is set to
smu
X
.OUTPUT_NORMAL
.
Example
smua.source.offfunc = smua.OUTPUT_DCVOLTS
Sets the normal output-off mode to
source 0 V when the output is
turned off for SMU channel A.
Also see
(on page 7-220)
(on page 7-221)
smuX.source.offlimitY
This attribute sets the limit (current or voltage) used when the source-measure unit (SMU) is in normal output-off
mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
Current: 1e-3 (1 mA)
Voltage: 40 (40 V)
Summary of Contents for 2657A