Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-344
2651A-901-01 Rev. A / March 2011
Details
The
trigger.timer[
N
].reset()
function resets the following attributes to their factory defaults:
•
trigger.timer[
N
].count
•
trigger.timer[
N
].delay
•
trigger.timer[
N
].delaylist
•
trigger.timer[
N
].passthrough
•
trigger.timer[
N
].stimulus
It also clears
trigger.timer[
N
].overrun
.
Also see
(on page 7-340)
(on page 7-343)
(on page 7-344)
trigger.timer[N].stimulus
This attribute specifies which event starts the timer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Trigger timer N reset
Not saved
0
Usage
eventID
= trigger.timer[
N
].stimulus
trigger.timer[
N
].stimulus =
eventID
eventID
The event that triggers the timer delay
N
A number representing a trigger timer number (1 to 4)
Details
The
eventID
parameter may be one of the trigger event IDs shown in the following table.
Trigger event IDs*
Event ID
Event description
smu
X
.trigger.SWEEPING_EVENT_ID
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
smu
X
.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
smu
X
.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smu
X
.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measure action