status.questionable.instrument.*
This attribute contains the questionable status instrument summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
2 (All bits set)
Usage
questionableRegister
= status.questionable.instrument.condition
questionableRegister
= status.questionable.instrument.enable
questionableRegister
= status.questionable.instrument.event
questionableRegister
= status.questionable.instrument.ntr
questionableRegister
= status.questionable.instrument.ptr
status.questionable.instrument.enable =
questionableRegister
status.questionable.instrument.ntr =
questionableRegister
status.questionable.instrument.ptr =
questionableRegister
questionableRegister
The questionable status instrument summary register's status. A zero (0)
indicates no bits set (also send 0 to clear all bits); the only valid value
other than zero (0) is 2
Details
These attributes are used to read or write to the questionable status instrument summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register bits are set.
The least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
(on page E-
Enable and transition registers
(on page E-18). The individual bits of this register have the following
meanings:
Bit Value
Description
B0
Not used
Not applicable
B1
status.questionable.instrument.SMUA
Set bit indicates one or more enabled bits in the
questionable status SMU A summary event
register is set.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2-B15
Not used
Not applicable.
In addition to the above constants,
questionableRegister
can be set to the numeric equivalent of the bit to
set.
Example
status.questionable.instrument.enable =
status.questionable.instrument.SMUA
Sets the SMUA bit of the
questionable status instrument
summary enable register using a
constant.