Section 2: General operation
Model 2651A High Power System SourceMeter® Instrument Reference Manual
2-30
2651A-901-01 Rev. A / March 2011
CAUTION
When using the I-Source as a sink, ALWAYS set V-Compliance to a level that is higher than the
external voltage level. Failure to do so could result in excessive current flow into the Model 2651A
(<102 mA) and incorrect measurements. See Compliance limit for details.
NOTE
The only exception to the compliance limit not being exceeded is the VLIMIT when operating as an
ISOURCE. To avoid excessive (and potentially destructive) currents from flowing, the VLIMIT will
source or sink up to 102 mA for ISOURCE ranges on or below 100 mA. For the ranges 1 A and
above, the maximum current allowed is the current source setting.
The sink operating limits are shown in General High Power System SourceMeter
®
Instrument power
equation in the
Configuring for measure only tests using the MODE key
In addition to being used for conventional source-measure operations, the Model 2651A can also be
used like a meter to measure current, voltage, resistance, or power.
To configure the Model 2651A as a V-meter, I-meter, ohm-meter, or watt-meter:
1. Press
the
MODE
key.
2. Turn the navigation wheel
to select the desired meter from the menu (
I-METER
,
V-METER
,
OHM-METER
, or
WATT-METER
) and then press the
ENTER
key.
3. The Model 2651A has now been configured to the selected meter.
To manually configure the settings, refer to the applicable topics:
•
V-meter and I-meter measurements
(on page 2-30)
•
(on page 2-31)
•
V-meter and I-meter measurements
To make V-meter and I-meter measurements without using the MODE key (such as when configuring
measure-only tests over the remote interface), follow the basic procedure below.
Perform the following steps to use the Model 2651A to measure voltage or current:
1. Select source-measure functions.
Voltmeter:
Press the
SRC
key to select the I-source, and press the
MEAS
key to select the
voltage measurement function.
I-meter (ammeter):
Press the
SRC
key to select the V-source, and press the
MEAS
key to select
the current measurement function.
2. Set source and compliance levels. Use the editing procedure provided in steps 1 and 2 of the
Front-panel source-measure procedure
(on page 2-24) to edit the source and compliance levels:
a. Select the lowest source range and set the source level to zero.
b. Set compliance to a level that is higher than the expected measurement.