Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-210
2651A-901-01 Rev. A / March 2011
Also see
(on page 3-6)
smuX.sense
This attribute contains the state of the remote/local sense mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall
Saved setups
smuX.SENSE_LOCAL
Usage
senseMode
= smu
X
.sense
smu
X
.sense =
senseMode
senseMode
The sense mode. Set to one of the following:
0
or
smu
X
.SENSE_LOCAL
: Selects local sense (2-wire).
1
or
smu
X
.SENSE_REMOTE
: Selects remote sense (4-wire).
3
or
smu
X
.SENSE_CALA
: Selects calibration sense mode.
X
SMU channel (for example,
smua.sense
applies to SMU Channel A).
Details
Source-measure operations are performed using either 2-wire local sense connections or 4-wire remote sense
connections. Writing to this attribute selects the sense mode.
The
smu
X
.SENSE_CALA
mode is only used for calibration and may only be selected when calibration is enabled.
The sense mode can be changed between local and remote while the output is on.
The calibration sense mode cannot be selected while the output is on.
Resetting the instrument selects the local sense mode.
Example
smua.sense = smua.SENSE_REMOTE
Selects remote sensing for SMU
Channel A.
Also see
(on page 2-43)