Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-133
localnode.description
This attribute stores a user-defined description of the instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
Instrument specific (see
Details
)
Usage
localnode.description =
description
description
= localnode.description
description
User-defined description of the instrument
Details
This attribute stores a string that contains a description of the instrument. This value appears on instrument's LXI
home page.
This attribute's default value contains
Keithley
MMMM
#
SSSSSSSS
, where:
MMMM
is the instrument's four-digit
model number, and #
SSSSSSSS
is the instrument's eight-digit serial number. You can change it to a value that
makes sense for your system.
When using this command from a remote node,
localnode
should be replaced with the node reference, for
example
node[5].description
.
Example
description = "System in Lab 05"
localnode.description = description
Set
description
equal to "System in Lab 05".
Set the LXI home page of this instrument to display
"System in Lab 05".
Also see
Using the web interface in the Model 2651A User's manual
localnode.linefreq
This attribute contains the line frequency setting used for NPLC calculations.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
60
Usage
frequency
= localnode.linefreq
localnode.linefreq =
frequency
frequency
An integer representing the instrument's detected or specified line frequency