2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-231
Details
When set to
smu
X
.SOURCE_HOLD
, this attribute disables pulse mode sweeps, holding the source
level for the remainder of the step.
When set to
smu
X
.SOURCE_IDLE
, this attribute enables pulse mode sweeps, setting the source level
to the programmed (idle) level at the end of the pulse.
Example
smua.trigger.endpulse.action = smua.SOURCE_IDLE
smua.trigger.endpulse.stimulus = trigger.timer[1].EVENT_ID
Configure the end pulse action to achieve a pulse and configure trigger timer 1 to control the end of pulse.
Also see
(on page 3-30)
smuX.trigger.endpulse.set()
This function sets the end pulse event detector to the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu
X
.trigger.endpulse.set()
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.set()
applies to SMU channel A)
Details
This function sets the end pulse event detector to the detected state.
The SMU automatically clears all the event detectors when the
smu
X
.trigger.initiate()
function is executed. Therefore,
smu
X
.trigger.endpulse.set()
should be called after the
sweep is initiated. If the event detectors are configured to clear automatically because the
smu
X
.trigger.autoclear
attribute is set to
smu
X
.ENABLE
, make sure that
smu
X
.trigger.endpulse.set()
is issued after the SMU has entered the trigger layer.
Also see
(on page 7-229)
(on page 3-30)