Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-230
2606B-901-01 Rev. B / May 2018
The overrun statuses of the event detectors are not automatically cleared when the SMU transitions
from the arm layer to the trigger layer.
The event detectors are always cleared when a sweep is initiated.
Also see
(on page 3-30)
smuX.trigger.count
This attribute sets the trigger count in the trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
1
Usage
triggerCount
= smu
X
.trigger.count
smu
X
.trigger.count =
triggerCount
triggerCount
The trigger count is the number of times the source-measure unit (SMU) will iterate
in the trigger layer for any given sweep
X
SMU channel (for example,
smua.trigger.count
applies to SMU channel A)
Details
During a sweep, the SMU iterates through the trigger layer of the trigger model this many times. After
performing this many iterations, the SMU returns to the arm layer.
If this count is set to zero (
0
), the SMU stays in the trigger model indefinitely until aborted.
Also see
(on page 3-30)
smuX.trigger.endpulse.action
This attribute enables or disables pulse mode sweeps.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
1 (smuX.SOURCE_HOLD)
Usage
pulseAction
= smu
X
.trigger.endpulse.action
smu
X
.trigger.endpulse.action =
pulseAction
pulseAction
The pulse mode setting; set to one of the following values (see Details for
definition):
0
or
smu
X
.SOURCE_IDLE
1
or
smu
X
.SOURCE_HOLD
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.action
applies to SMU channel A)