smuX.measure.nplc
This command sets the integration aperture for measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
1.0
Usage
nplc
= smu
X
.measure.nplc
smu
X
.measure.nplc =
nplc
nplc
The integration aperture; set from 0.001 to 25
X
Source-measure unit (SMU) channel (for example,
smua.measure.nplc
applies
to SMU channel A)
Details
This attribute controls the integration aperture for the analog-to-digital converter (ADC).
The integration aperture is based on the number of power line cycles (NPLC), where 1 PLC for 60 Hz
is 16.67 ms (1/60) and 1 PLC for 50 Hz is 20 ms (1/50).
Example
smua.measure.nplc = 0.5
Sets the integration time for SMU channel A to 0.5.