smuX.measure.filter.count
X = SMU channel (a or b)
Attribute
Number of measured readings to yield one filtered measurement.
Default
1
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
count = smuX.measure.filter.count
-- Reads filter count.
smuX.measure.filter.count = count
-- Writes filter count.
count
Set filter count from 1 to 100
.
Remarks
• This attribute is the number of measurements that will be performed to yield one filtered
measurement.
• The
reset
function sets the filter count to 1.
Details
See
Also see
Example
Sets filter count for SMU A:
smua.measure.filter.count = 10
smuX.measure.filter.enable
X = SMU channel (a or b)
Attribute
Enables/disables filtered measurements.
Default
smuX.FILTER_OFF
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
filter = smuX.measure.filter.enable --Reads on/off state of the filter.
smuX.measure.filter.enable = filter --Writes on/off state of the filter.
filter
The filter status.
Set
filter
to one of the following values:
0
or
smuX.FILTER_OFF
Disables the filter.
1
or
smuX.FILTER_ON
Enables the filter.
Remarks
• This attribute enables or disables the filter.
• The
reset
function disables the filter.
Details
See
Also see
Example
Enable the filter for SMU A:
smua.measure.filter.enable = smua.FILTER_ON