2600AS-901-01 Rev. B / September 2008
Return to
19-115
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
smuX.measure.delay
X= SMU channel (a or b)
Attribute
Controls the measurement delay.
Default
smuX.DELAY_OFF
(2601A/2602A/2611A/2612A)
smuX.DELAY_AUTO
(2635A/2636A)
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
mdelay = smuX.measure.delay
-- Reads the measure delay.
smuX.measure.delay = mdelay
-- Writes the measure delay.
mdelay
The measurement delay value.
Set
mdelay
to one of the following values:
0
or
smuX.DELAY_OFF
No delay.
-1
or
smuX.DELAY_AUTO
Automatic delay value.
user_value
Set user delay value.
Remarks
• This attribute allows for additional settling time before taking a measurement.
• The
smuX.DELAY_AUTO
setting causes a current range-dependent delay to be inserted when a
current measurement is requested. This happens when a current measurement command is
executed, when the measure action is being performed in a sweep, or after changing ranges
during an auto-ranged measurement.
• If
smuX.measure.count
is greater than 1, the measurement delay is only inserted before the
first measurement.
•
mdelay
can be set to a specific user-defined value that sets the delay that is used regardless of
range.
Also see
,
Example
Sets the measurement delay to 10mS:
smua.measure.delay = 0.010
smuX.measure.delayfactor
X= SMU channel (a or b)
Attribute
This attribute is a multiplier to the delays used when
smuX.delay
is set to
smuX.DELAY_AUTO
.
Default
1
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
delayfactor = smuX.measure.delayfactor -- Reads the delay factor.
smuX.measure.delayfactor = delayfactor -- Writes the delay factor.
delayfactor
The delay factor multiplier.
Remarks
• The delay factor is only applied when
smuX.measure.delay = smuX.DELAY_AUTO.
• The default value is 1.0.
• This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic
delay.
• This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the
automatic delay.
• Setting this attribute to zero disables delays, even when
smuX.measure.delay = smuX.DELAY_AUTO
.
Also see
Example
Increase the automatic delay by 2 times:
smuX.measure.delayfactor = 2.0