Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
smuX.contact.r
X = SMU channel (a or b)
Function
Measure contact resistance.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
rhi, rlo = smuX.contact.r()
rhi
The measured contact resistance on the high/sense high side.
rlo
The measured contact resistance on the low/sense low side.
Remarks
• Attempting to perform a contact check measurement when any of the following conditions exist
will generate an error:
• Output is off in High-Z mode.
• Current limit set to less than 1mA.
Details
See
for connections.
Also see
smuX.contact.speed
X = SMU channel (a or b)
Attribute
The speed setting for contact check measurements.
Default
smuX.CONTACT_FAST
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
speed_opt = smuX.contact.speed
-- Reads speed setting.
smuX.contact.speed = speed_opt
-- Writes speed setting.
speed_opt
The speed setting.
Set
speed_opt
to one of the following:
0
or
smuX.CONTACT_FAST
1
or
smuX.CONTACT_MEDIUM
2
or
smuX.CONTACT_SLOW
Remarks
• This setting controls the aperture of measurements made for contact check. It does not affect the
smuX.measure.nplc
aperture setting.
• The speed setting can have a dramatic effect on the accuracy of the measurement, as reflected
in the specifications.
Example
Set contact check measurements on SMU A for higher accuracy:
smua.contact.speed = smua.CONTACT_SLOW