Grounded Phoenix-to-Triax Cable Adapter 

PA-1004 Rev. A / June 2009 

 

Connections 

This section identifies the connections that are made within the cable adapter and how to connect it between the 
instruments and fixtures. 

WARNING 

Do not touch connectors when this assembly is connected; possible exposure to hazardous 
voltages could result in personal injury or death. 

 
 

You must provide an additional safety barrier to prevent the possibility of electrical shock if 
this cable adapter is used with an instrument capable of sourcing voltages greater the 42 V. 

 
 

No interlock features are provided by this device; properly make all connections before 
turning on the power to any  instrumentation connected to this assembly. 

Table 1

 describes the connection pathways between the rear-panel terminal blocks of the System SourceMeter 

instruments and Models 8101-4TRX and LR-8028 test fixtures. 

Table 1: Connector pathway description

 

Connector 

Outer shield 

Inner shield 

Center conductor 

Triax HI 

LO SMU Pin 1 

Guard SMU Pin 3 

HI SMU Pin 4 

Triax LO 

LO SMU Pin 1 

LO SMU Pin 1 

LO SMU Pin 1 

Banana GND 

N/A 

N/A 

LO SMU Pin 1 

 

The banana jack, which is connected to the chassis through a frequency variable resistor (FVR), can be used as a 
common signal ground point for input and output LO signals. 

Figure 2 

shows the connector schematic for both the 

triax connectors and the low-noise chassis ground banana jack.  