6-2
Range, Filter, and Math
Model 2520 User’s Manual
Range
Measurement ranges
The selected measurement range affects the accuracy of the laser diode voltage measure-
ments, photodiode detector current measurements, as well as the maximum signal that can
be measured.
Laser diode voltage ranges
summarizes laser diode voltage ranges, resolution, and maximum readings.
Photodiode detector current ranges
lists the available current measurement ranges, resolution values, and maximum
readings for each Model 2520 photodiode detector.
NOTE
The current measurement range can be individually set for each detector.
Table 6-1
Laser diode voltage measurement ranges
Voltage
range
Resolution
Maximum
reading
5V
0.33mV
±5.25V
10V
0.66mV
±10.5V
Table 6-2
Photodiode current measurement ranges
Current
range
Resolution
Maximum
reading
10mA
0.7
µ
A
±10.5mA
20mA
1.4
µ
A
±21mA
50mA
3.4
µ
A
±52.5mA
100mA
6.8
µ
A
±105mA