4-14
Troubleshooting
Model 2520 Pulsed Laser Diode Test System Service Manual
Power supply checks
Power supply problems can be checked out using
for circuit theory on the power supply.
Table 4-2
Power supply checks
Step
Item/
component*
Required condition
Remarks
1
2
3
4
5
6
7
8
9
10
11
12
Line fuse
Line power
TP1
TP2
TP14
TP15
TP16
TP20
TP21
TP22
TP23
TP32
Check continuity.
Plugged into live receptacle, power on.
+25V, ±10%
-25V, ±10%
+25V, ±5%
+12V, ±5%
+5V, ±5%
+5V, ±5%
-5V, ±5%
+12V, ±5%
-12V, ±5%
+3.3V, ±5%
Remove to check.
Check for correct power-up sequence.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
* Test points (TP) are marked on circuit boards.
Summary of Contents for 2520
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 30: ......
Page 31: ...2 Calibration...
Page 68: ...2 38 Calibration Model 2520 Pulsed Laser Diode Test System Service Manual...
Page 69: ...3 RoutineMaintenance...
Page 72: ...3 4 Routine Maintenance Model 2520 Pulsed Laser Diode Test System Service Manual...
Page 73: ...4 Troubleshooting...
Page 90: ...4 18 Troubleshooting Model 2520 Pulsed Laser Diode Test System Service Manual...
Page 91: ...5 Disassembly...
Page 98: ...5 8 Disassembly Model 2520 Pulsed Laser Diode Test System Service Manual...
Page 99: ...6 ReplaceableParts...
Page 116: ...6 18 Replaceable Parts Model 2520 Pulsed Laser Diode Test System Service Manual...
Page 117: ...A Specifications...
Page 121: ...B CalibrationReference...
Page 132: ...B 12 Calibration Reference Model 2520 Pulsed Laser Diode Test System Service Manual...
Page 136: ......