Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-246
2461-901-01 A/November 2015
Details
Setting the pulse width to zero (0) seconds asserts the trigger indefinitely. To release the trigger line,
use
trigger.digout[
N
].release()
.
Example
digio.line[4].mode = digio.MODE_TRIGGER_OUT
trigger.digout[4].pulsewidth = 20e-6
Sets the pulse width for trigger line 4 to
20
µ
s.
Also see
(on page 8-50)
(on page 8-52)
Digital I/O port configuration
(on page 3-87)
(on page 8-244)
(on page 8-246)
trigger.digout[N].release()
This function releases an indefinite length or latched trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.digout[
N
].release()
N
Digital I/O trigger line (1 to 6)
Details
Releases a trigger that was asserted with an indefinite pulsewidth time. It also releases a trigger that
was latched in response to receiving a synchronous mode trigger. Only the specified trigger line is
affected.
Example
digio.line[4].mode = digio.MODE_TRIGGER_OUT
trigger.digout[4].release()
Releases digital I/O trigger line 4.
Also see
(on page 8-50)
Digital I/O port configuration
(on page 3-87)
(on page 8-244)
(on page 8-245)