Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-20
2461-901-01 A/November 2015
Example
:CALC:VOLT:MATH:FORM MXB
:CALC:VOLT:MATH:MMF 0.80
:CALC:VOLT:MATH:MBF 50
:CALC:VOLT:MATH:STAT ON
Set the math function for voltage measurements to
mx+b.
Set the scale factor for voltage measurements to 0.80.
Set the offset factor to 50.
Enable the math function.
Also see
Calculations that you can apply to measurements
(on page 3-82)
:CALCulate[1]:<function>:MATH:MBFactor
(on page 6-20)
:CALCulate[1]:<function>:MATH:MMFactor
(on page 6-22)
:CALCulate[1]:<function>:MATH:PERCent
(on page 6-23)
:CALCulate[1]:<function>:MATH:STATe
(on page 6-24)
:CALCulate[1]:<function>:MATH:MBFactor
This command specifies the offset, b, for the y = mx + b operation.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0
Usage
:CALCulate[1]:<function>:MATH:MBFactor <n>
:CALCulate[1]:<function>:MATH:MBFactor DEFault
:CALCulate[1]:<function>:MATH:MBFactor MINimum
:CALCulate[1]:<function>:MATH:MBFactor MAXimum
:CALCulate[1]:<function>:MATH:MBFactor?
:CALCulate[1]:<function>:MATH:MBFactor? DEFault
:CALCulate[1]:<function>:MATH:MBFactor? MINimum
:CALCulate[1]:<function>:MATH:MBFactor? MAXimum
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
The digitize function:
•
Digitize current:
DIGitize:CURRent
•
Digitize voltage:
DIGitize:VOLTage
<n>
The offset for the y = mx + b operation; the valid range is
−
1e12 to
+
1e12