2.
Select the DMM DC volts function, and enable auto-ranging.
3.
Enable the Model 2303 DVM input.
4.
Turn on the Model 2303 source output.
5.
Verify digital voltmeter input accuracy for each of the voltages listed in Table 1-8. For
each test point:
• Set the voltage to the indicated value as measured by the digital multimeter.
• Set compliance to 3A.
• Allow the reading to settle.
• Verify that the Model 2303 voltage reading is within the limits given in the table.
* As measured by digital multimeter. See procedure.
Table 1-8
Digital voltmeter input accuracy limits
Model 2303 output
voltage setting*
Digital voltmeter input reading limits
(1 Year, 18°C-28°C)
14.00V
13.990 to 14.010V
3.00V
2.996V to 3.004V
深圳德标仪器
135-1095-0799
Summary of Contents for 2303
Page 12: ...1 Performance Verification 135 1095 0799...
Page 26: ...2 Calibration 135 1095 0799...
Page 48: ...3 Disassembly 135 1095 0799...
Page 53: ...3 6 Disassembly 135 1095 0799...
Page 76: ...A Specifications 135 1095 0799...
Page 81: ...A 7 Specifications 135 1095 0799...
Page 82: ...B CalibrationReference 135 1095 0799...
Page 90: ...C CalibrationPrograms 135 1095 0799...