3-30
Pulse Current Measurements
Pulse current measurements
The following command sequence will return the average of 10 peak pulse current
measurements:
Battery channel (#1)
DISP:CHAN 1
‘ Sets active channel - battery.
SENS:CURR:RANG 5
‘ Select 5A range.
VOLT 15
‘ Set output voltage to 15V.
CURR 0.75
‘ Set current limit to 750mA.
OUTP ON
‘ Turn output on.
SENS:PCUR:SYNC ON
‘ Enable trigger synchronization.
SENS:PCUR:AVER 10
‘ Set average count to 10.
SENS:PCUR:SYNC:TLEV:ONE 0.1
‘ Set trigger level to 100mA for 1A trigger
level range.
SENS:PCUR:SYNC:TLEV:RANG 0.5
‘ Select the 1A trigger level range.
SENS:PCUR:TIME:AUTO
‘ Set integration times automatically.
SENS:PCUR:MODE HIGH
‘ Configure to measure peak pulse.
SENS:FUNC “PCUR”
‘ Select pulse current function.
READ?
‘ Trigger 10 measurement conversions and
‘ return the average of those 10 conver-
‘ sions. The average of the 10 conversions
‘ is displayed on the front panel. Each of
‘ the ten conversion syncs to the rising
‘ edge.
Charger channel (#2)
DISP:CHAN 2
‘ Sets active channel - charger.
SENS2:CURR:RANG 5
‘ Select 5A range.
SOUR2:VOLT 15
‘ Set output voltage to 15V.
SOUR2:CURR 0.75
‘ Set current limit to 750mA.
OUTP2 ON
‘ Turn output on.
SENS2:PCUR:SYNC ON
‘ Enable trigger synchronization.
SENS2:PCUR:AVER 10
‘ Set average count to 10.
SENS2:PCUR:SYNC:TLEV 0.1
‘ Set trigger level to 100mA.
SENS2:PCUR:TIME:HIGH 600e-3
‘ Set integration high time to 600ms.
SENS2:PCUR:SYNC:DEL 50e-3
‘ Set trigger delay to 50msec.
SENS2:PCUR:MODE HIGH
‘ Configure to measure peak pulse (trigger
‘ on rising edge).
SENS2:FUNC “PCUR”
‘ Select pulse current function.
READ2?
‘ Trigger 10 measurement conversions and
‘ return the average of those 10 conver-
‘ sions. The average of the 10 conversions
‘ is displayed on the front panel. Each of
‘ the ten conversion syncs to the rising
‘ edge.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...