Pulse Current Measurements 3-13
SCPI programming — pulse current measurements
The commands for pulse current measurements are summarized in
(a listing
following the table contains specific command notes).
“Programming examples” on page 3-29
demonstrate how to use these commands.
Table 3-2
SCPI commands — pulse current measurements
Command
Description
Default
SENSe[1]
SENSe subsystem for Channel #1 (battery channel):
:FUNCtion “PCURrent”
Select pulse current measurement function.
VOLT
:PCURrent
Pulse current configuration:
:AVERage <NRf>
Specify average count:
1–100 (pulse current measurements) or
1–5000 (pulse current digitization).
1
:MODE <name>
Select measurement mode; HIGH, LOW or AVERage.
HIGH
:TIME
Set integration times:
:AUTO
Integration times set automatically.
:HIGH <NRf>
Specify integration time (in sec) for high pulse measurements; 33.33e-6
to 0.8333.
3.333e-5
:LOW <NRf>
Specify integration time (in sec) for low pulse measurements; 33.33e-6 to
0.8333.
3.333e-5
:AVERage <NRf>
Specify integration time (in sec) for average pulse measurements; 33.33e-
6 to 0.8333.
3.333e-5
:DIGitize <NRf>
Specify integration time (in sec) for digitizing or burst measurements (B10
firmware and later); 33.33e-6 to 0.8333.
3.333e-5
:SYNChronize
Pulse detection triggering:
[:STATe]
Send ON to select pulse current measurements or
OFF to select pulse current digitization.
ON
:TLEVel
Trigger level:
[:AMP] <NRf>
Set trigger level (in amps) for 5A range: 0.0–5.0
0.0
:ONE <NRf>
Set trigger level (in amps) for 1A range: 0.0–1.0
0.0
:MILLiamp <NRf>
Set trigger level (in amps) for 100mA range: 0.0–0.1
0.0
:HALFamp <NRf>
Model 2306-PJ only — set trigger level (in amps) for 500mA range: 0-
500mA.
0.0
:HUNDred <NRf>
Model 2306-PJ only — set trigger level (in amps) for 100mA range: 0-
100mA.
0.0
:TEN <NRf>
Model 2306-PJ only — set trigger level (in amps) for 10mA range: 0-
10mA.
0.0
:RANGe <NRf>
Model 2306, 2306-VS or 2306-PJ when on 5A current range.
Set trigger level range (100mA, 1A, or 5A). The parameter
<NRf> sent with this command causes the trigger to be set with
the trigger level setting of MILL, ONE, or AMP. Queries
receive responses of 0.1, 1.0, or 5.0 accordingly. In other
words, if a value of 2.0A is sent with the command, a value of
5A will be returned as a response to a query.
5A
SENSe[1]
:PCURrent
:SYNChronize
:TLEVel
:RANGe <NRf>
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...