D-2
Calibration Program
Introduction
This appendix includes a calibration program written in BASIC to help you in calibrating the
Model 2306. Refer to Section 14 for more details on calibration procedures, equipment, and
connections. Appendix C covers calibration commands in detail.
Computer hardware requirements
The following computer hardware is required to run the calibration programs:
•
IBM PC compatible computer.
•
Keithley KPC-488.2 or KPC-488.2AT, or CEC PC-488 IEEE-488 interface for the
computer.
•
Two shielded IEEE-488 bus cables (Keithley Model 7007).
Software requirements
To use the calibration program, you will need the following computer software:
•
Microsoft QBasic (supplied with MS-DOS 5.0 or later).
•
MS-DOS version 5.0 or later.
•
HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and CEC
interface cards listed above).
Calibration equipment
The following calibration equipment is required:
•
Keithley Model 2001 Digital Multimeter
•
4
Ω
, 0.1%, 100W resistor
•
3k
Ω
, 0.1%, 0.25W resistor
Refer to Section 14 for detailed equipment specifications as well as details on test connec-
tions.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...