Performance Verification 13-17
1.
Select the DMM DC volts function, and enable auto-ranging.
2.
Select Channel #1 by pressing the
▲
and
▲
keys to toggle between Channel #1 and
Channel #2, and enable the Model 2306 DVM input.
3.
Turn on the Model 2306 source output.
4.
Verify digital voltmeter input accuracy for each of the voltages listed in
. For
each test point:
• Set the voltage to the indicated value as measured by the digital multimeter.
• Set current compliance to 0.25A.
• Allow the reading to settle.
• Verify that the Model 2306 voltage reading is within the limits given in the table.
1.
Repeat steps 1 through 5 for Channel #2 (use the
▲
and
▲
keys to toggle between
Channel #1 and Channel #2).
Table 13-8
Digital voltmeter input accuracy limits
Model 2306 output
Voltage setting
1
Digital voltmeter input reading limits
(2 Years, 18°C-28°C)
5.00V
4.995 to 5.005V
10.00V
9.992 to 10.008V
14.00V
13.990 to 14.010V
1
As measured by digital multimeter.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...