Series 2280 Precision DC Power Supply Reference Manual
Section 7: SCPI command reference
077085503 / March 2019
7-123
If you forget the password, you can reset the password to the default. On the front panel, press
MENU
. Under System, select Manage. Select
Password Reset.
In the pop-up window, select
LXI
to
reset the password.
Example
:SYST:PASS:NEW "admin","123456"
Change the password of the instrument to 123456
Also see
(on page 7-122)
(on page 7-123)
:SYSTem:PASSword:UNLock
This command enables a remote interface.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
admin
Usage
:SYSTem:PASSword:UNLock <password>
<password>
The password for remote mode
Details
When a remote interface is locked, sending this command with the correct password will release it.
Sending this command with wrong password will cause the error, 509, "System password not
correct."
Example
:SYST:PASS:UNL "admin"
Unlock the present interface used to send this
command.
Also see
(on page 7-122)
(on page 7-122)