![Keithley 2260B Series Quick Start Manual Download Page 17](http://html2.mh-extra.com/html/keithley/2260b-series/2260b-series_quick-start-manual_5096856017.webp)
2260B Quick Start Guide
14
2.
The OVP setting will be displayed on the top
and the OCP setting (or OFF) will be displayed
on the bottom.
OVP Level
3.
Use the Voltage knob to set the
OVP level.
Range
10%~110% of rated
output voltage.
OCP Level
4.
Use the Current knob to set the
OCP level.
Range
10%~110% of rated
output current, OFF.
5.
Press OVP/OCP again to exit. The
OVP/OCP indicator will turn off.
Power switch trip
Set F-95 (Power switch trip) to 1 (to
disable the power switch trip) or to
0 (to enable the power switch trip)
and save.
F-95
1 (Disable) or 0 (Enable)
Clear OVP/OCP
protection
The OVP or OCP protection can be
cleared after it has been tripped by
holding the OVP/OCP button for 2
seconds.
(Only applicable when the power
switch trip setting is disabled
[F-95 = 1])
(hold)