Basic Operation
Setting OVP/OCP Levels
Background
For most models the OVP level has a selectable
range of approximately* 10% to 110% of the
rated output voltage. Likewise the OCP level
for most models has a selectable range of
approximately* 10%~ 110% of the rated output
current. The OVP and OCP level is set to the
maximum by default. The OCP level can also
be turned off.
*Note that the
actual
setting range differs for
each model.
When one of the protection measures are on,
ALM is shown on the panel display. By
default, the power switch will turn off when
any of the protection levels are tripped.
ALM
V
A
Before setting the OVP or OCP level:
Ensure the load is not connected.
Ensure the output is set to off.
Setting Ranges
2260B (360W)
30-36
80-13
250-4
800-1
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
3.6-39.6
1.35-14.85 0.45-4.95
0.144-1.584
2260B (720W)
30-72
80-27
250-9
800-2