D-2
Model 182 Emulation Commands
The Model 2182 can be configured to accept device-dependent commands of the Keithley
Model 182 Sensitive Digital Voltmeter. The commands for controlling the Model 2182 with the
182 language are provided in
. For details on Model 182 operation, refer to the
Model 182 Instruction Manual.
Since the architecture of the Model 2182 differs from that of the Model 182, some commands
are different and cannot be used. Be sure to refer to the notes at the end of the table for
information on command restrictions.
CAUTION
The 182 language is intended to be used only over the IEEE-488 bus. Using
front panel controls in conjunction with this language may cause erratic
operation. In this case, results cannot be guaranteed.
Table D-1
Model 182 device-dependent command summary
Mode
Command
Description
Note
Display ASCII String
A0
Restore display to normal
1
A1,string
Display string (up to 12 characters)
Display Resolution
B0
5
½
-digit resolution
B1
6
½
-digit resolution
B2
3
½
-digit resolution
B3
4
½
-digit resolution
Calibration
None
C (Calibration) commands not supported
2
Filter Damping
D0
Configure filter damping off (same as P2)
D1
Configure filter damping on (same as P3)
Reading Source
F0
Latest reading from A/D converter
F1
One reading from buffer
F2
All readings in buffer
F3
Maximum value in buffer
F4
Minimum value in buffer
Reading Format
G0
Reading only
3
G1
Reading with prefix
G2
Reading with buffer location
G3
Reading with buffer location and prefix
Immediate Trigger
H0
Initiate manual trigger
4
Buffer Configuration
I0
Disable buffer
5
I1,value
Buffer on, value = buffer size
Analog Output
J0
Disable analog output relative
6
Relative
J1
Enable analog output relative using last reading as Rel
value
J2,value
Enable analog output relative using value (±1e–9 to ±120)
J3
Enable analog output relative
EOI, Bus Hold-off
K0
Enable EOI, enable bus hold-off on X
K1
Disable EOI, enable bus hold-off on X
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...