Common Commands
12-13
Description
Use the *SRE command to program the Service Request Enable Register. Send this command
with the decimal equivalent of the binary value that determines the desired state (0 or 1) of each
bit in the register. This register is cleared on power-up.
This enable register is used along with the Status Byte Register to generate service requests
(SRQ). With a bit in the Service Request Enable Register set, an SRQ occurs when the
corresponding bit in the Status Byte Register is set by an appropriate event. For more
information on register structure, see the information presented earlier in this section.
The Service Request Enable Register is shown in
. Notice that the decimal weight
of each bit is included in the illustration. The sum of the decimal weights of the bits that you
wish to set is the value that is sent with the *SRE command. For example, to set the ESB and
MAV bits of the Service Request Enable Register, send the following command:
*SRE 48
Where:
ESB (bit B5)
= Decimal
32
MAV (bit B4) = Decimal
16
<NRf>
=
48
The contents of the Service Request Enable Register can be read using the *SRE? query
command.
Figure 12-3
Service request enable register
B7
B6
B5
B4
B3
B2
B1
B0
(2 )
4
(2 )
5
16
32
0/1
0/1
Bit Position
Event
Decimal Weighting
Value
Value: 1 = Enable Service
Request Event
0 = Disable (Mask)
Service Request Event
Events : OSB = Operation Summary Bit
ESB = Event Summary Bit
MAV = Message Available
QSB = Questionable Summary Bit
EAV = Error Available
MSB = Measurement Summary Bit
OSB
ESB
QSB
MAV
EAV
MSB
(2 )
7
128
(2 )
3
8
(2 )
2
4
(2 )
0
1
0/1
0/1
0/1
0/1
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...