Device action
The primary device action is a measurement. However, the device action block could include
the following additional actions (refer to
Figure 7-2
Device action
•
Filtering
— If the repeating filter is enabled, the instrument samples the specified
number of reading conversions to yield single filtered reading. Only one reading
conversion is performed if the filter is disabled, or after the specified number of reading
conversions for a moving average filter is reached. After a reading (Rdg) is procured,
operation proceeds to Hold.
•
Hold
— The Hold feature is used to screen out reading anomalies. When enabled, the
user selects a window and count for Hold. In general, when a reading is outside the
window it is rejected; operation loops back to be beginning of the Device Action as
shown in
. The hold count specifies how many readings have to be within the
window before it is accepted. See “
” for operation details. After
a Hold Reading is acquired, operation proceeds to Channel Closure.
•
Channel closure
— When stepping or scanning, the last device action is channel
control. Using the hold feature provides an auto settling time for switching relays. Each
open/close transition will restart the hold process and a reading for each channel will not
occur until the relay settles.
Output trigger
After the device action, an output trigger occurs and is available at the rear panel Trigger Link
connector. This trigger can be used to trigger another instrument to perform an operation (e.g.,
select the next channel for an external scan).
Rdg
Hold
Chan
Filter
DEVICE ACTION
From Delay block
of Figure 7-1
To Output Trigger
block of Figure 7-1
Triggering
7-5
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...