4.
When the unit is done calibrating, it will display the following prompt:
REMOVE INPUT
5.
Remove the low-thermal short, and press ENTER. During this phase, the
CALIBRATING message will be displayed.
DC volts calibration
1.
After the front panel short and open procedure, the unit will prompt you for the first DC
voltage:
10.000000 V
2.
Connect the calibrator to the CHANNEL 1 input of the Model 2182 using the low-
thermal cable, as shown in
. Wait two minutes to allow for thermal equilibrium
before proceeding. Model 2107 cable connections are summarized in
Figure 2-2
Connections for DC volts calibration
Table 2-3
Input cable color codes
Input connection
Wire color
Channel 1 HI
Red
Channel 1 LO
Black
Channel 2 HI
Green
Channel 2 LO
White
!
CHANNEL 1
CHANNEL 2
Model 2182
2107 Low
Thermal Cable
OUTPUT LO
Red (CH1)
Green (CH2)
OUTPUT HI
Black (CH1)
White (CH2)
DC Voltage Calibrator
2-8
Calibration
Summary of Contents for 2182
Page 10: ......
Page 14: ...1 Performance Verification Performance Verification...
Page 25: ...1 12 Performance Verification...
Page 26: ...2 Calibration Calibration...
Page 43: ...2 18 Calibration...
Page 44: ...3 Routine Maintenance RoutineMain tenance...
Page 47: ...3 4 Routine Maintenance...
Page 48: ...4 Troubleshooting Troubleshoot ing...
Page 64: ...5 Disassembly Disassembly...
Page 73: ...5 10 Disassembly...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ...6 Replaceable Parts Replaceable Parts...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ...A Specifications Specifica tions...
Page 96: ...B CalibrationReference Calibration Reference...
Page 109: ...B 14 Calibration Reference...
Page 110: ...C CalibrationProgram Calibration Program...
Page 115: ...C 6 Calibration Program...
Page 119: ......