Warm-up period
Allow the Model 2182 to warm up for at least 2-1/2 hours before conducting the verification
procedures. If the instrument has been subjected to temperature extremes (those outside the
ranges stated above), allow additional time for the instrument’s internal temperature to stabilize.
Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified
temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the
manufacturer.
Line power
The Model 2182 requires a nominal line voltage of 100, 120, 220, or 240V and a line
frequency of 50 to 60Hz or 400Hz. Verification tests must be performed with the correct line
voltage. If necessary, change the line voltage setting as covered in
.
Recommended test equipment
summarizes recommended verification equipment. You can use alternate equip-
ment as long as that equipment has specifications at least four times better than corresponding
Model 2182 specifications. Keep in mind, however, that test equipment accuracy will add to the
uncertainty of each measurement.
Table 1-1
Recommended verification equipment
Description
Manufacturer/Model
Specifications*
DC Calibrator
Fluke 5700A
100mV:
±
14ppm
1V:
±
7ppm
10V:
±
5ppm
100V:
±
7ppm
Low Thermal Divider
100:1 division: ±5ppm
Thermocouple Calibrator
Omega CL523
Type J: -200 to 700°C
Type K, N : -200 to 1300°C
Type N: -200 to 400°C
Type E: -200 to 1000°C
Type R, S: 0 to 1700°C
Type B: 400 to 1700°C
Digital Multimeter
Keithley 2000
1V:
±
32ppm
Low Thermal Cable
Keithley 2107
Low Thermal Cable
Keithley 1506
BNC-to-dual banana jack cable
*DC calibrator specifications shown include total 90-day 23°C ±5°C uncertainty at specified output.
Model 262 error includes a short-term transfer error of
±
5ppm at 23°C
±
1°C. Digital multimeter
specification is 90-day, 23°C ±5°C full-range value.
Performance Verification
1-3
Summary of Contents for 2182
Page 10: ......
Page 14: ...1 Performance Verification Performance Verification...
Page 25: ...1 12 Performance Verification...
Page 26: ...2 Calibration Calibration...
Page 43: ...2 18 Calibration...
Page 44: ...3 Routine Maintenance RoutineMain tenance...
Page 47: ...3 4 Routine Maintenance...
Page 48: ...4 Troubleshooting Troubleshoot ing...
Page 64: ...5 Disassembly Disassembly...
Page 73: ...5 10 Disassembly...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ...6 Replaceable Parts Replaceable Parts...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ...A Specifications Specifica tions...
Page 96: ...B CalibrationReference Calibration Reference...
Page 109: ...B 14 Calibration Reference...
Page 110: ...C CalibrationProgram Calibration Program...
Page 115: ...C 6 Calibration Program...
Page 119: ......