RTD temperature
1.
Connect the precision decade resistance box (listed in
INPUT and SENSE jacks using 4-wire connections. (See
connecting scheme.)
2.
Configure the Model 2010 temperature function for °C units and RTD temperature
sensor (
α
=0.00385) as follows:
A.
Press SHIFT then SENSOR, and note the unit displays the temperature units:
UNITS: C.
B.
Press ENTER, and note the unit displays the sensor type: SENS: TCOUPLE.
C.
Using the cursor and range keys, set the display as follows: SENS: 4W-RTD.
D.
Press ENTER, and note the unit displays: TYPE: PT100.
E.
Using the cursor and range keys, set the unit for the following display: TYPE:
PT385.
F.
Press ENTER to complete the temperature configuration process.
3.
Select the temperature function by pressing the TEMP key.
4.
Set the decade resistance box to each of the values shown in
, and verify that
the temperature readings are within the required limits.
Table 1-8
Four-wire RTD temperature verification reading limits
Applied resistance*
Reading limits (°C)
(1 year, 18°C–28°C)
22.80
Ω
100.00
Ω
313.59
Ω
-190.14 to -189.86°C
-0.08 to +0.08°C
599.86 to 600.14°C
* Based on a = 0.00385. See text.
Performance Verification
1-13
Summary of Contents for 2010
Page 14: ...1 Performance Verification 1 Performance Verification...
Page 28: ...2 Calibration 2 Calibration...
Page 53: ...3 Routine Maintenance 3 Routine Maintenance...
Page 56: ...4 Troubleshooting 4 Troubleshoot ing...
Page 74: ...5 Disassembly 5 Disassembly...
Page 83: ...CONFIDENTIAL PROPRIETARY...
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ...6 Replaceable Parts 6 Replaceable Parts...
Page 98: ...CONFIDENTIAL PROPRIETARY...
Page 99: ...CONFIDENTIAL PROPRIETARY...
Page 100: ......
Page 101: ...A Specifications A Specifications...
Page 107: ...B Calibration CommandReference B Calibration CommandRef erence...