2000-903-01 Rev. C / October 2007
Return to
Section Topics
1-37
Model 2000 Digital Multimeter Quick Start Guide
Section 1: Introduction
Single-trigger ?
Trigger measurement
GET
G0
G1
G2
G3
G4
G5
G6
G7
G8
G9
Get instrument configuration
(function, range, speed,
trigger)
Get SRQ mask
Get calibration input prompt
(generates error message
51)
Get user-defined message
(replies with 16 space
characters)
Get calibration status (replies
with “1000”)
Get IAB status (input F/R,
autorange on/off, offset on/
off)
Get YW status (suffix
enabled/disabled, terminator
selection)
Get error status
Get instrument ID
(replies with
“FLUKE,8842A,0,EMUL”)
Get status of JKM commands
(temp. units; TC type and
junction)
String = 1jkm
GET (cont.)
G10
G11
Get closed channel number
String = 10nn
Where: nn =
00 (all open)
01 through 10 (closed
channel)
Get simulated reference
junction temperature
String = xx.xxx (in °C)
Note: G2 valid only in
calibration mode.
PUT
P0
P1
P4
Put instrument configuration
(function, range, speed,
trigger)
Put SRQ mask
Set simulated junction
temperature; 0 to 50 (°C)
Default temperature = 23°C
PUT format
N <value> P0
N <value> P1
N <value> P4
Thermocoupl
e type
K0
K1
K2
Type J thermocouple
(default)
Type K thermocouple
Type T thermocouple
Temperature
unit
J0
J1
J2
°C (default)
K
°F
Table 1-17
Models 8840A/8842A device-dependent command summary (cont.)
Mode
Command
Description