background image

 

2

 

WARNING

 

This high voltage probe is designed to prevent accidental shock to 
the operator when properly used. This operating note must be read  
and fully understood prior to using the probe.

 

Specifications

 

Input resistance:

Approx. 1000m

 

 

Division ratio:

1000:1

Maximum working voltage:

40KV DC or Peak AC, 28KV RMS AC

Accuracy:

DC volts:

±1%(1KV to 20KV), ±2%(20KV to 40KV)

AC volts:

Typically 5% at 60Hz

Temperature Coefficient:

Less than 200ppm/ºC

Operating temperature:

0ºC to +50ºC

Storage temperature:

-20ºC to -70ºC

Cable length:

1 meter

 

Safety precautions

 

This high voltage probe must only be used by personnel who are trained, experienced, or otherwise 
qualified to recognize hazardous situations and who are trained in the safety precautions that are 
necessary to avoid possible injury when using such a device.

Do not work alone when working with high voltage circuits.

For your own safety, inspect the probes for cracks and frayed or broken leads before each use. If defects 
are noted, DO NOT use the probe.

Hands, shoes, floor, and work bench must be dry. Avoid making measurements under humid, damp, or 
other environmental conditions that might affect the safety of the measurement situation.

If possible, always turn the high voltage source off before connecting or disconnecting the probe.

The probe body should be kept clean and free of any conductive contamination. Refer to the section on 
cleaning.

 

Operation

 

1.

Connect the plugs to the volts (HI) and corn (LO) input terminals of your voltmeter.

2.

Select the desired voltmeter function and range: do not use autoranging.

3.

Whenever possible, turn the high voltage source off before making any connections.

4.

Connect the divider probe common lead (alligator clip) to a good earth ground or reliable chassis 
ground.

 

WARNING

 

Do not attempt to take measurements from sources where the 
chassis or return lead is not grounded.