8.  Fit  ports  either  side  of  the  IRmadilloNA

’s install point to allow clean-in-place or flushing 

with a suitable cleaning fluid. 

1.3.  Checking for warnings and checking the probe is clean 

When you have connected to the powered-on IRmadilloNA the KeitSpec application should 
load  automatically  (within  3  minutes).  It  is  important  to  confirm  that  the  system  is  working 
without  errors:  information  on  IRmadilloNA  warnings  and  errors  can  be  found  in 

DOC0893 

KeitSpec Software User Manual

, section 1, 

Getting Started

Once the system is set up without errors, the probe tip must be cleaned, and a background 
spectrum must be taken before insertion into the vessel or pipe. If the probe is guaranteed 
clean, the factory background scan shipped with the instrument 

– and saved to the desktop of 

the controller 

– can be used. Cleanliness of the probe tip is important to performance; if you 

suspect fouling or scale build-up, perform a clean appropriate to your environment.  

1.3.1.  Cleaning instructions 

Rinse  the  ATR  element  with  a  suitable  solvent.  The  choice  of  solvent  will  depend  on  your 
process. You may fully immerse the ATR in solvent if needed. The following are suitable: 

  Water 

  Acetone 

  Alcohols 

  Surfactant solutions 

 

Alkanes (hexane, cyclohexane etc…) 

  Ethers 

Once exposed to a suitable solvent, wipe the ATR with a clean cloth. The metal tube of the 
probe can also be cleaned by wiping with a cloth. 

For  hard-to-remove  fouling  (such  as  biofilms  or  dried-on  chemicals)  Keit  recommends  an 
oxidative acid clean. First, clean the ATR using the methods outlined above, then perform the 
following: Souriau in-line plug 

  If you have purchased one, place the PTFE sample cell on the end of the probe and 

fill  with  ~4ml  of  1 M  nitric  acid solution.  If you  have  not  purchased  one  then  please 
contact [email protected] for further information. 

  Leave to soak for 1 hr 

  Carefully remove the acid using a pipette and fill the sample cell with water 

  Remove the water using a pipette 

  Remove the sample cell and rinse the ATR again with water 

  Carefully wipe the ATR with a cloth 

It  may  be  necessary  to  repeat  the  cleaning  steps  again  if  the  contaminant  has  only  been 
partially removed. 

In some cases, it may be necessary to perform an alkali clean on the probe. First clean the 
ATR using the methods outline above, then perform the following: 

  Place the PTFE sample cell on the end of the probe and fill with ~4ml of 1 M sodium 

hydroxide solution 

  Leave to soak for 1 hr 

  Carefully remove the alkali using a pipette and fill the sample cell with water 

  Remove the water using a pipette 

Summary of Contents for IRmadillo DOC1049R

Page 1: ...DOC1049R 22 February 2024 12 07 DOC1049R IRmadilloNA User Manual Hardware Model Number ASM0627 09 N Cx O Gx D3x www keit co uk support keit co uk...

Page 2: ...nd 7 2 Maintenance 8 2 1 Spectrum health 8 2 1 1 Use at high temperature 8 2 1 2 Emitter ageing 9 2 2 Remote health check 9 2 3 Cleaning the fibre optic data cable connectors 9 2 3 1 Cleaning the Sour...

Page 3: ...e used The output from this purge gas generator should have a minimum quality according to ISO8573 1 2010 of Class 1 2 1 dust water oil or better Keit can recommend suitable generators that achieve th...

Page 4: ...If the purge tubing is disconnected the blanking plugs must be fitted to the inlet connections on the probe and body to maintain the IP rating and exclude dust The IRmadilloNA requires a stabilisation...

Page 5: ...reduced diameter section of the pipe The minimum flow velocity should be 1 5 m s 5 ft s 4 Install the IRmadilloNA so that the probe tip is fully in the flow We recommend putting the probe tip as close...

Page 6: ...d The following are suitable Water Acetone Alcohols Surfactant solutions Alkanes hexane cyclohexane etc Ethers Once exposed to a suitable solvent wipe the ATR with a clean cloth The metal tube of the...

Page 7: ...of interest within the spectrum is between 800 and 1800 cm 1 This section of the absorbance plot should maintain a relatively flat line with no consistently repeating peaks Cleaning Mode can be used t...

Page 8: ...other background must be taken before any spectra can be measured Figure 6 Negative peaks indicating the background was taken when the probe was unclean or not fully purged 1 4 Taking a background FTI...

Page 9: ...ment 2 1 1 Use at high temperature Figure 7 Probe is heated causing the spectrum to tilt In Figure 7 the spectrum is tilted with positive absorbance at lower wavenumbers and negative absorbance at hig...

Page 10: ...perform an annual examination of your system s health data remotely to determine if any servicing needs to be carried out This includes review of internal diagnostics for warnings and errors component...

Page 11: ...rease the probability of fracture Prot gez la fibre des contraintes et des courbures serr es rayon de courbure 30 mm interdit Les courbures serr es provoquent une contrainte la surface du verre et aug...

Page 12: ...en The pen fits over each of the two male fibre ferrules and is aligned by the transparent cover Push and click once on each ferrule Figure 12 Using the fibre cleaning pen to the Souriau plug Once bot...

Page 13: ...Page 12 DOC1049R 22 February 2024 12 07 www keit co uk Back to Table of Contents Figure 13 Reconnecting the Souriau connectors Locknut Align keyways...

Page 14: ...ions can cause damage to the instrument If in doubt please consult with Keit Le stockage du spectrom tre IRmadilloNA non aliment et non purg dans des conditions inad quates peut endommager l instrumen...

Page 15: ...t force we recommend to tighten the thumbscrew until resistance is felt 1 turn For volatile substances it is recommended to keep the stopper cap in place during use Figure 14 Sample cell fitted to IRm...

Page 16: ...s is an alternative to installing the IRmadilloNA into a main pipeline It is fixed in place using a tri clamp that locates over the rear of the flow cell and two semi circular collars that fit around...

Page 17: ...ency Annual 4 2 2 Low flow cell ASM1222 The low flow cell that can be secured to the end of the probe for continuous analysis of your process The flow cell is equipped with two 1 16 NPT fittings to at...

Page 18: ...ll have not perished They should show no signs of swelling tearing or degradation Also inspects the clamp and C pieces for any signs of degradation cracks or damage If the clamp and or the C pieces sh...

Page 19: ...r as a bench top instrument for offline sampling Angled at 35 the probe can be positioned into beakers or other vessels for ease of use It can also be configured for two heights to give added flexibil...

Page 20: ...meter over such that the mounting holes are pointed upward the instrument should be disconnected from power coms and purge to avoid damage to the cables or twisting Position the inclined stand over th...

Reviews: