To install the inclined stand to the instrument, carefully roll the spectrometer over such that
the mounting holes are pointed upward, the instrument should be disconnected from power
coms and purge to avoid damage to the cables or twisting.
Position the inclined stand over the mounting points such that the mounting plate lines up with
the instrument. The spectrometer is located with 4 screws through the corners of the stand
into the mounting holes on the spectrometer.
Return the instrument to the correct orientation and connect the purge, the power cable, and
communications cable/s.
5.8. Ingold DN40
A version of the IRmadillo spectrometer probe is available to allow connection to reaction
vessels or process pipework fitter with an Ingold DN40 port. The probe tube is fitted with a
Ingold DN40 nut that can be fitted to compatible ports. For technical drawings, contact Keit.
It is important to follow these instructions completely to ensure safe use of the Ingold probe.
Figure 32: IRmadillo fitted with an Ingold probe
5.8.1. Safety
WARNING
The O-ring and nut form a seal with the process pipework and as
such are safety critical to prevent an accidental discharge of the
process media. Keit recommends regular inspection of the port in
line with local safety procedures.
CAUTION
The spectrometer weighs ~18kg. The probe must not be used to
support the instrument on process pipework. Doing so risks
excess stress on the piping system.